Two Tier Calibration of a Vector Network Analyzer is a technique that allows two calibration error sets to be combined into one
img4
Two Tier Calibration of a Vector Network
What is Two Tier Calibration?
calibration error set alone.
This technique solves VNA calibration problems that are not easily solved by one
lib ti t l
img5
img6img7
coaxial port on one side and a probed port on the other side.
Another example would be a device with a different probe pitch on each port.
The dilemma is finding a“Thru”calibration standard to use for the
y p
you will run into a dilemma to complete the calibration with one set of calibration standards.
The most common example of a DUT with mixed ports is one with a i l t id d b d t th th id
The dilemma is finding a Thru”calibration standard to use for the
calibration of the VNA between such ports.Short,Open,and Load standards for the VNA calibration of the two ports are rather simple.The“Thru”is the issue.
When you have a device under test that has mixed port connections
img8img9
me to solve?
What sort of problem will Two Tier Calibration help me to solve?
img10
Thru,assuming a suitable thru exists at all.This is like the Chicken and the Egg;Which came first?
img11
q)
Effectively the Two Tier Calibration yields the S-Parameters of the Thru,thus solving the dilemma.
And you’ll see that Two Tier Calibration is a simpler procedure.
For the example of a DUT with a coaxial port and a probed port,you could complete a 2-port VNA calibration if you had the S-Parameters of a second probe to use as the calibration Thru.
You face the same dilemma,how to measure the S-Parameters of the
complete the calibration?
Can’t I just use S-Parameters of the Thru to
complete the calibration?
Even though Cascade Microtech provides S-Parameters for probes we
manufacture,your best VNA calibration should include your cable to probe connection,and your VNA sweep setup(number of points,frequencies,IF bandwidth,etc).
Even though Cascade Microtech provides S Parameters for probes we
img12
img13
VNA with the additional probe calibration.
–The result is you have the VNA calibrated now to your two desired reference planes,coax on one side and probe tip on the other side.
which will combine the existing two port calibration already in the
img14img15
reference planes,coax on one side and probe tip on the other side.
Two Tier Calibration Procedure
a probed port on the other side,here is the procedure:
–From the front panel of the VNA,complete a full two port coaxial calibration to the ends of the two VNA test port cables using coaxial calibration standards or an electronic calibration module.
For the common example of a DUT with a coaxial port on one side and a probed port on the other side here is the procedure:
–Connect one VNA test port cable to the probe you are using to contact the probed side of your DUT.
–Within the WinCal XE calibration menu select“Second Tier”.
–And then complete a 1-Port calibration to the tip of the probe using Short,Open,and Load calibration standards.
–Finish by sending the probe calibration error terms to the VNA,which will combine the existing two port calibration already in the
img16
p og ess e y t oug t e data e e d esu t s a S parameters of the probe are determined.It works great!”
S12 S21.The last product is pulled apart by square root with a
Engineering:“Measurement of three standards,Short,Open,and Load at the tip of the probe provides three(complex)equations with three unknowns(complex):S11,S22,and
S12*S21 The last product is pulled apart by square root with a
img17
It works great!
Engineering:“Measurement of three standards Short Open
Marketing:“It works great!”
through the data.The end result is all 4 S-
How does it work?
img18
img19img20
WinCal XE as shown below.
img21img22img23
•After finishing the two port coaxial calibration of the VNA from the instrument front panel,proceed with
WinCal XE as shown below
img24
Step by Step for This Example
img25
complete you are done with the calibration!See how simple it is.
img26img27img28
•Finish by sending the probe calibration error terms to the VNA,which will combine the existing two port calibration already in the VNA with the additional probe calibration.After this step is complete you are done with the calibration!See how simple it is
img29
Step by Step for This Example
img30img31img32img33img34
WinCal XE
More about this Superior Microwave Tool
Visit www.cascademicrotech.com/wincal for more information
FREE DEMO VERSION
–Unlimited data viewing,manipulation and display
FREE TRIAL VERSION
–Fully functional for 30 days Academic discount available